Article 2018

Flip-angle based ratiometric approach for pulsed CEST-MRI pH imaging

{Several molecules have been exploited for developing MRI pH sensors based on the chemical exchange saturation transfer (CEST) technique. A ratiometric approach, based on the saturation of two exchanging pools at the same saturation power, or by varying the saturation power levels on the same pool, is usually needed to rule out the concentration term from the pH measurement. However, all these methods have been demonstrated by using a continuous wave saturation scheme that limits its translation to clinical scanners. This study shows a new ratiometric CEST-MRI pH-mapping approach based on a pulsed CEST saturation scheme for a radiographic contrast agent (iodixanol) possessing a single chemical exchange site. This approach is based on the ratio of the CEST contrast effects at two different flip angles combinations (180\mbox{$^\circ$}/360\mbox{$^\circ$} and 180\mbox{$^\circ$}/720\mbox{$^\circ$}), keeping constant the mean irradiation RF power (Bavg power). The proposed ratiometric approach index is concentration independent and it showed good pH sensitivity and accuracy in the physiological range between 6.0 and 7.4.}

Author(s): Arena, C and Irrera, P and Consolino, L and Colombo Serra, S and Zaiss, M and Longo, DL
Journal: {Journal of Magnetic Resonance}
Volume: 287
Pages: 1--9
Year: 2018
Bibtex Type: Article (article)
DOI: 10.1016/j.jmr.2017.12.007
Electronic Archiving: grant_archive

BibTex

@article{ArenaICCZL2017,
  title = {{Flip-angle based ratiometric approach for pulsed CEST-MRI pH imaging}},
  journal = {{Journal of Magnetic Resonance}},
  abstract = {{Several molecules have been exploited for developing MRI pH sensors based on the chemical exchange saturation transfer (CEST) technique. A ratiometric approach, based on the saturation of two exchanging pools at the same saturation power, or by varying the saturation power levels on the same pool, is usually needed to rule out the concentration term from the pH measurement. However, all these methods have been demonstrated by using a continuous wave saturation scheme that limits its translation to clinical scanners. This study shows a new ratiometric CEST-MRI pH-mapping approach based on a pulsed CEST saturation scheme for a radiographic contrast agent (iodixanol) possessing a single chemical exchange site. This approach is based on the ratio of the CEST contrast effects at two different flip angles combinations (180\mbox{$^\circ$}/360\mbox{$^\circ$} and 180\mbox{$^\circ$}/720\mbox{$^\circ$}), keeping constant the mean irradiation RF power (Bavg power). The proposed ratiometric approach index is concentration independent and it showed good pH sensitivity and accuracy in the physiological range between 6.0 and 7.4.}},
  volume = {287},
  pages = {1--9},
  year = {2018},
  slug = {arenaicczl2017},
  author = {Arena, C and Irrera, P and Consolino, L and Colombo Serra, S and Zaiss, M and Longo, DL}
}