Bloch-Siegert flip angle calibration for phosphorus at the human brain at 9.4 T using ISIS localization
{Correct calibration of the transmit field B1+ is crucial to achieve optimal SNR. However, fast and robust B1+ calibration is difficult for X-nuclei due to the low signal sensitivity. In this work, we proposed a fast B1+ calibration method based on the Bloch-Siegert shift and single voxel ISIS localization. With the proposed sequence, the B1+ calibration can be done in less than 5 min for the human brain at B0 \textequals 9.4 T.}
@misc{item_3248092, title = {{Bloch-Siegert flip angle calibration for phosphorus at the human brain at 9.4 T using ISIS localization}}, booktitle = {{2020 ISMRM \& SMRT Virtual Conference \& Exhibition}}, abstract = {{Correct calibration of the transmit field B1+ is crucial to achieve optimal SNR. However, fast and robust B1+ calibration is difficult for X-nuclei due to the low signal sensitivity. In this work, we proposed a fast B1+ calibration method based on the Bloch-Siegert shift and single voxel ISIS localization. With the proposed sequence, the B1+ calibration can be done in less than 5 min for the human brain at B0 \textequals 9.4 T.}}, year = {2020}, slug = {item_3248092}, author = {Ruhm, L and Dorst, J and Avdievich, N and Henning, A} }