B0 shimming for X-nuclei imaging at 7 T using 23Na B0 maps
{B0 inhomogeneities are a major challenge at ultra-high magnetic field strengths (B0 $\geq$ 7 T), as they can lead to strong image artifacts. In this work, a B0 shimming routine that is based on 23Na B0 maps was implemented for applications in X-nuclei imaging where no 1H MRI data can be acquired and therefore, no vendor-provided shimming routines can be used. The proposed 23Na ConsTru shimming routine showed an improvement in B0 homogeneity comparable to the vendor-provided GRE Brain shimming routine both in phantom and in vivo measurements using 23Na B0 map acquisition times less than 1 minute.}
@misc{item_3054577, title = {{B0 shimming for X-nuclei imaging at 7 T using 23Na B0 maps}}, booktitle = {{ISMRM 27th Annual Meeting \& Exhibition}}, abstract = {{B0 inhomogeneities are a major challenge at ultra-high magnetic field strengths (B0 $\geq$ 7 T), as they can lead to strong image artifacts. In this work, a B0 shimming routine that is based on 23Na B0 maps was implemented for applications in X-nuclei imaging where no 1H MRI data can be acquired and therefore, no vendor-provided shimming routines can be used. The proposed 23Na ConsTru shimming routine showed an improvement in B0 homogeneity comparable to the vendor-provided GRE Brain shimming routine both in phantom and in vivo measurements using 23Na B0 map acquisition times less than 1 minute.}}, year = {2019}, slug = {item_3054577}, author = {Gast, LV and Henning, A and Hensel, B and Uder, M and Nagel, AM} }