Homogenous Excitation in Whole Brain CEST: Combination of Snapshot CEST and Multiple Interleaved Mode Saturation
{To perform Chemical Exchange Saturation Transfer MRI of the whole brain a homogeneous saturation and fast readout are required. To achieve a fast and robust 3D acquisition a spiral-centric-reordered GRE readout was used. In addition, a Multiple Interleaved Mode Saturation scheme was applied to mitigate B1+-inhomogeneity effects of the CEST saturation. Combination of these two methods allows acquiring a homogenous CEST contrast in a volume of approximately 220x220x45mm3 within an acquisition time of 7 min 26s.}
@misc{item_3054598, title = {{Homogenous Excitation in Whole Brain CEST: Combination of Snapshot CEST and Multiple Interleaved Mode Saturation}}, booktitle = {{ISMRM 27th Annual Meeting \& Exhibition}}, abstract = {{To perform Chemical Exchange Saturation Transfer MRI of the whole brain a homogeneous saturation and fast readout are required. To achieve a fast and robust 3D acquisition a spiral-centric-reordered GRE readout was used. In addition, a Multiple Interleaved Mode Saturation scheme was applied to mitigate B1+-inhomogeneity effects of the CEST saturation. Combination of these two methods allows acquiring a homogenous CEST contrast in a volume of approximately 220x220x45mm3 within an acquisition time of 7 min 26s.}}, year = {2019}, slug = {item_3054598}, author = {Liebert, A and Zaiss, M and Gumbrecht, R and Liebig, P and Schmitt, B and Laun, F and Doerfler, A and Uder, M and Nagel, A} }